(1) instrument features
Thermo Scientific TVA2020C toxic volatile gas analyzer is the only toxic volatile gas analyzer using both flame ionization (FID) and photoionization (PID) dual detector technology. It has the ability to detect both organic and inorganic compounds at the same time. Compared with other instruments of the same volume, it can provide more comprehensive gas coverage. Compared to earlier models, the TVA2020C weighs 21 percent less and is more compact than most single FID instruments. This instrument has the following features:2) technical parameters of tva-2020c toxic volatile gas analyzer
Main performance indicators
Technical parameters
The accuracy of
The maximum value of FID was 1.0~10000ppm, plus or minus 10% or plus or minus 0.1 PPM
PID from 0.5 to 500ppm, plus or minus 20% or plus or minus 0.5 PPM of the reading, taking the maximum value
repetitive
At 500ppm methane, 2%
When PID 100ppm isobutene, 1%
Linear range
FID 1.0 ~ 30000 PPM methane
PID 0.5~2000ppm isobutylene
The response time
The FID was injected into 10000ppm methane, and the time to reach the final value was less than 3.5s
PID was injected into 500ppm isobutylene, and the time to reach the final value was less than 3.5s
Sampling flow
At the entrance of the sampling probe, it is generally 1L/min
The battery
When 0 ℃, can work continuously at least 10 hours. The battery is fully charged for less than 10 hours
Hydrogen bottle working hours
Start with cylinder pressure of 15.3Mpa(2,200Psi) and work continuously for more than 10 hours
weigh
Single FID detector 4.17kg
Dual detector 4.26 kg
Working environment temperature
-10℃~45℃
Minimum detection limit
The minimum detection limit is calculated by the standard deviation of noise between seven peaks
PID 0.5 PPM isobutylene
FID 0.5 PPM methane
life
PID routine cleaning, over 2000 hours
More than 5,000 hours of FID
Data storage interval
Automatic mode 1 time/second or 1 time /999 minutes, user can choose
VOC or FE mode, 2~30 seconds, user optional
Working relative humidity
15%~95%
voltage
220VAC 50/60Hz
Duration of field work
With spare gas cylinders and external power supply, it can work continuously for more than 50 hours
Storage function
The machine has data storage function, which can save no less than 9000 data
Name: | |